Skip to main content
OPEN NOW
0s

0% of today's working hours elapsed

Office Hours (IST)

Mon – Fri9:00 AM – 6:00 PM
Saturday9:00 AM – 3:00 PM
SundayClosed
Track Repair →
Doble M4000 Transformer Test Kit Repair - Power Plant, Raichur – Industrial Electronics Repair | Synchronics
Power PlantEquipmentDoble Engineering₹26.90 Lakh Saved

Doble M4000 Transformer Test Kit Repair - Power Plant, Raichur

69h
Turnaround
77%
Cost saved
Raichur, Karnataka
Location
Doble M4000-HV-CH3
Part No.

Fault Analysis

The Problem

HV measurement channel zero-offset drift from +2 mV to +18 mV, exceeding manufacturer tolerance band

Leakage current sensor preamplifier sensitivity degradation of 3.2%, causing non-linear core-loss measurement error

Core-loss measurement reading 8% low (0.8% vs expected 0.89%), failing transformer acceptance criteria

AD8066 JFET-input op-amp input offset voltage drift beyond specification limits in HV measurement buffer stage

Film capacitor aging in leakage current sensor gain-setting network, degrading preamplifier frequency response

Money Saved

Cost Savings Breakdown

What this customer saved by repairing instead of buying new

OEM New Unit
₹35.00 Lakh
Would have paid
Repair Cost
₹8.10 Lakh
Actually paid
You Saved
₹26.90 Lakh
77% savings
Repair CostOEM Price
₹8.10 Lakh paid₹26.90 Lakh saved

* Govt. taxes extra as applicable · Shipping charges extra as applicable on the repair quote.

This Repair's Journey

From Inquiry to Delivery

Completed in 69 hours

STEP 01

Inquiry Received

Equipment details logged, initial assessment scheduled

STEP 02

Diagnosis & Evaluation

Component-level fault analysis with detailed report

STEP 03

Repair Work

Expert component-level repair using OEM-grade parts

STEP 04

Testing & QC

Full-load testing under operating conditions, QA sign-off

STEP 05

Delivery & Warranty

Dispatched with test report and 6-month warranty

69h
Total Turnaround
6 months
Warranty Included

Technical Details

Our Repair Approach

1

High-voltage amplifier offset diagnostics and isolation

Oscilloscope measurement of the HV amplifier output confirmed +18 mV DC offset deviation from the manufacturer specification band of -2 to +2 mV, isolating the fault to the input buffer stage op-amp.

2

Leakage current sensor sensitivity verification and component tracing

NIST-traceable 100 mA reference current injection testing revealed 3.2% sensitivity loss in the preamplifier stage, traced to degraded OPA2134 op-amp and aged precision resistor network.

3

Precision micro-soldering removal and replacement of failed AD8066 op-amp

Controlled removal and replacement of the AD8066 JFET-input operational amplifier in the HV measurement input buffer, preserving all neighboring high-precision analog filtering components and PCB integrity.

4

Leakage current sensor resistor network and coupling capacitor replacement

Surgical replacement of five 10 kOhm 1% thin-film resistors in the matched gain-setting array and the polypropylene coupling capacitor bank using precision component matching and impedance verification.

5

Full offset nulling and preamplifier sensitivity re-calibration

Executed M4000 internal calibration routine to null the HV channel offset to -0.3 mV (within spec) and re-calibrated leakage current sensor sensitivity against a NIST-certified 100 mA reference source.

6

Side-by-side validation testing and commissioning report certification

Comparative testing against a reference Doble M4000 unit confirmed core-loss measurement accuracy within 0.05% and generated IEC 60076-1 compliance certification for power plant transformer acceptance.

Mission Accomplished

The Result

The repaired Doble M4000 kit was returned to the power plant within 69 hours, enabling immediate transformer acceptance testing and commissioning completion on schedule. Synchronics' repair cost of INR 810,000 delivered 77% cost savings versus the OEM replacement quote of INR 3,500,000, preserving INR 2,690,000 in client budget and eliminating the 2-week procurement lead-time risk.

69h
Turnaround
77%
Cost Saved
6 mo
Warranty
Free
No-fix, no charge

Have similar equipment that needs repair?

Free diagnosis · No-fix, no-charge · 6-month warranty · Pan-India pickup

Trusted Since 1994 · ISO Certified · 32+ Years

Get Your Equipment
Back in Action — Fast

Send us your faulty industrial electronics. Our engineers will diagnose it and provide a detailed quote within 24 hours — with no obligation to proceed.

Free Evaluation

No obligation

48hr Turnaround

Fastest in India

🛡

6-Month Warranty

Every repair

No Fix, No Charge

Zero risk policy

📦