Skip to main content
OPEN NOW
0s

0% of today's working hours elapsed

Office Hours (IST)

Mon – Fri9:00 AM – 6:00 PM
Saturday9:00 AM – 3:00 PM
SundayClosed
Track Repair →
Omicron CMC356 Protection Relay Test Set - Utilities, Visakhapatnam – Industrial Electronics Repair | Synchronics
IndustrialEquipmentOmicron Electronics₹22.70 Lakh Saved

Omicron CMC356 Protection Relay Test Set - Utilities, Visakhapatnam

52h
Turnaround
71%
Cost saved
Visakhapatnam, Andhra Pradesh
Location
CMC356-6
Part No.

Fault Analysis

The Problem

SD Card Read Error - file allocation table (FAT) corruption preventing access to all stored test sequences and calibration data

Firmware Checksum Failure - bootloader and application firmware corrupted in Winbond W25Q128 SPI NOR flash memory

POST Sequence Abort - processor unable to complete power-on self-test due to incomplete firmware image verification

Brownout Protection Failure - three failed backup capacitors (3300 µF 16V) in power supply rail unable to sustain voltage during shutdown

microSD Card Controller IC Failure - Samsung semiconductor address management logic damage preventing read/write operations on storage media

Money Saved

Cost Savings Breakdown

What this customer saved by repairing instead of buying new

OEM New Unit
₹32.00 Lakh
Would have paid
Repair Cost
₹9.30 Lakh
Actually paid
You Saved
₹22.70 Lakh
71% savings
Repair CostOEM Price
₹9.30 Lakh paid₹22.70 Lakh saved

* Govt. taxes extra as applicable · Shipping charges extra as applicable on the repair quote.

This Repair's Journey

From Inquiry to Delivery

Completed in 52 hours

STEP 01

Inquiry Received

Equipment details logged, initial assessment scheduled

STEP 02

Diagnosis & Evaluation

Component-level fault analysis with detailed report

STEP 03

Repair Work

Expert component-level repair using OEM-grade parts

STEP 04

Testing & QC

Full-load testing under operating conditions, QA sign-off

STEP 05

Delivery & Warranty

Dispatched with test report and 6-month warranty

52h
Total Turnaround
6 months
Warranty Included

Technical Details

Our Repair Approach

1

Firmware Extraction and Reflash Operation

Used specialized Omicron firmware programmer with JTAG boundary-scan interface to extract factory firmware binary from EEProm backup, performed sector-by-sector erase of corrupted Winbond W25Q128 SPI NOR flash, and reflashed with CRC validation and timestamp verification.

2

Corrupted SD Card Diagnosis and Replacement

Identified Samsung semiconductor controller IC failure in the microSD card, recovered 94% of test files (2847 of 3021 records) using specialized flash memory extraction and file carving algorithms, and replaced the card with OEM-equivalent Kingston 32GB Class 10 module.

3

Power Supply Capacitor Bank Replacement

Desoldered and replaced three failed 3300 µF 16V electrolytic capacitors in the backup power supply rail with upgraded 4700 µF 16V specification units to prevent future brownout failures during unplanned power loss events.

4

Firmware Integrity Validation and Bootloader Testing

Performed complete POST sequence validation with CRC checksum verification, confirmed proper bootloader execution, and validated all memory regions (code, data, calibration constants) against Omicron reference images.

5

Test Module Functional Verification

Executed comprehensive calibration testing across all 12 test modules (voltage, current, power, frequency, impedance, phase angle, harmonic, frequency sweep, transient, synchronization) using Omicron reference relay simulator and comparison with factory baseline values.

6

Recovered Data Restoration and System Acceptance Testing

Validated and reloaded 2847 recovered test records into the repaired CMC356-6, performed full file system integrity check on the new SD card, and completed final acceptance testing with customer's test engineer witnessing all module functions and data accessibility.

Mission Accomplished

The Result

The Omicron CMC356-6 was fully restored to operational status in 52 hours, enabling the utility to resume calibration testing across all 47 substations without further delay. Cost savings of INR 2.27 million (71% reduction vs. OEM replacement cost of INR 3.2 million) allowed the utility to reinvest funds into additional test equipment while recovering 18 months of archived test data valued at approximately INR 450,000 in compliance documentation and trending analysis.

52h
Turnaround
71%
Cost Saved
6 mo
Warranty
Free
No-fix, no charge

Have similar equipment that needs repair?

Free diagnosis · No-fix, no-charge · 6-month warranty · Pan-India pickup

Trusted Since 1994 · ISO Certified · 32+ Years

Get Your Equipment
Back in Action — Fast

Send us your faulty industrial electronics. Our engineers will diagnose it and provide a detailed quote within 24 hours — with no obligation to proceed.

Free Evaluation

No obligation

48hr Turnaround

Fastest in India

🛡

6-Month Warranty

Every repair

No Fix, No Charge

Zero risk policy

📦

Pan-India Pickup

Doorstep service